Measuring the Complete-arch Distortion of an Optical Dental Impression
Published on: May 30, 2019
Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A
Adolfo Di Fiore1, Lorenzo Graiff1, Gianpaolo Savio2
1Department of Neurosciences, Section of Prosthetic and Digital Dentistry, University of Padova, 35100 Padova, Italy.
Abstract:
Background: We compare the accuracy of new intraoral scanners (IOSs) in full-arch digital implant impressions. Methods: A master model with six scan bodies was milled in poly(methyl methacrylate), measured by using a coordinate measuring machine, and scanned 15 times with four IOSs: PrimeScan, Medit i500, Vatech EZ scan, and iTero. The software was developed to identify the position points on each scan body. The 3D position and distance analysis were performed. Results: The average and ± standard deviation of the 3D position analysis was 29 μm ± 6 μm for PrimeScan, 39 μm ± 6 μm for iTero, 48 μm ± 18 μm for Mediti500, and 118 μm ± 24 μm for Vatech EZ scan (p < 0.05). Conclusions: All IOSs are able to make a digital complete implant impression in vitro according to the average misfit value reported in literature (150 μm); however, the 3D distance analysis showed that only the Primescan and iTero presented negligible systematic error sources.