An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Identifying and overcoming limitations with in situ calibration beads for quantitative ultrasound
Jenna Cario1, Andres Coila1, Yuning Zhao1
1Department of Electrical and Computer Engineering, Beckman Institute for Advanced Science and Technology, Univerity of Illinois at Urbana-Champaign, Urbana, Illinois 61801, USA.
Abstract:
Ensuring the consistency of spectral-based quantitative ultrasound estimates in vivo necessitates accounting for diffraction, system effects, and propagation losses encountered in the tissue. Accounting for diffraction and system effects is typically achieved through planar reflector or reference phantom methods; however, neither of these is able to account for the tissue losses present in vivo between the ultrasound probe and the region of interest. In previous work, the feasibility of small titanium beads as in situ calibration targets (0.5-2 mm in diameter) was investigated. In this study, the importance of bead size for the calibration signal, the role of multiple echoes coming from the calibration bead, and sampling of the bead signal laterally through beam translation were examined. This work demonstrates that although the titanium beads naturally produce multiple reverberant echoes, time-windowing of the first echo provides the smoothest calibration spectrum for backscatter coefficient calculation. When translating the beam across the bead, the amplitude of the echo decreases rapidly as the beam moves across and past the bead. Therefore, to obtain consistent calibration signals from the bead, lateral interpolation is needed to approximate signals coming from the center of the bead with respect to the beam.