Technical note: Characterization and practical applications of a novel plastic scintillator for online dosimetry for
Yannick Poirier1,2, Junliang Xu1, Sina Mossahebi1
1Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland, USA.
Purpose:
Although flash radiation therapy (FLASH-RT) is a promising novel technique that has the potential to achieve a better therapeutic ratio between tumor control and normal tissue complications, the ultrahigh pulsed dose rates (UHPDR) mean that experimental dosimetry is very challenging. There is a need for real-time dosimeters in the development and implementation of FLASH-RT. In this work, we characterize a novel plastic scintillator capable of temporal resolution short enough (2.5 ms) to resolve individual pulses.
Methods:
We characterized a novel plastic dosimeter for use in a linac converter to deliver 16-MeV electrons at 100-Gy/s UHPDR average dose rates. The linearity and reproducibility were established by comparing relative measurements with a pinpoint ionization chamber placed at 10-cm water-equivalent depth where the electrometer is not saturated by the high dose per pulse. The accuracy was established by comparing the plastic scintillator dose measurements with EBT-XD Gafchromic radiochromic films, the current reference dosimeter for UHPDR. Finally, the plastic scintillator was compared against EBT-XD films for online dosimetry of two in vitro experiments performed at UHPDR.
Results:
Relative ion chamber measurements were linear with plastic scintillator response within ≤1% over 4-20 Gy and pulse frequencies (18-180 Hz). When characterized under reference conditions with NIST-traceability, the plastic scintillator maintained its dose response under UHPDR conditions and agreed with EBT-XD film dose measurements within 4% under reference conditions and 6% for experimental online dosimetry.
Conclusion:
The plastic scintillator shows a linear and reproducible response and is able to accurately measure the radiation absorbed dose delivered by 16-MeV electrons at UHPDR. The dose is measured accurately in real time with a greater level of precision than that achieved with a radiochromic film.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Photoluminescence: Applications
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET