Determining Intraosseous Needle Placement Using Point-of-Care Ultrasound in a Swine (Sus scrofa) Model
Adrianna I Kyle1, Jonathan D Auten1, Gregory J Zarow2
1Clinical Investigation Department, Combat Trauma Research Group, Naval Medical Center Portsmouth, Portsmouth, VA 23708, USA.
Military Medicine
|April 27, 2022
Summary
Point-of-care ultrasound with color power Doppler did not reliably confirm intraosseous (IO) line placement in swine models. However, experienced sonographers showed improved accuracy in identifying malpositioned IO catheters.
Area of Science:
- Emergency Medicine
- Ultrasound Technology
- Surgical Procedures
Background:
- Intraosseous (IO) access is vital for rapid resuscitation when peripheral IV access fails.
- Misplacement of IO lines can lead to serious complications like extravasation and tissue necrosis.
- Current research on ultrasound for IO placement confirmation is limited, especially in models relevant to young adults.
Purpose of the Study:
- To evaluate the effectiveness of bedside ultrasound in confirming IO needle placement.
- To assess ultrasound's utility in a live animal model with bone density similar to young adult patients.
- To investigate the impact of sonographer experience on accuracy.
Main Methods:
- A randomized, blinded prospective study using 36 adult swine.
- IO devices were placed in humeri, with 53 correctly and 19 incorrectly positioned.
- Four emergency physicians assessed placement using color power Doppler (CPD) ultrasound, blinded to needle location.
Main Results:
- Overall sensitivity for confirming IO placement was 72%, with a specificity of 79%.
- Sonographers with over 4 years of experience demonstrated higher specificity (86%) in identifying incorrect placements.
- Accuracy remained consistent between initial and final assessments.
Conclusions:
- Point-of-care ultrasound with CPD did not reliably confirm IO line placement in this swine model.
- Greater sonographer experience correlated with improved accuracy in identifying malpositioned IO catheters.
- Further research is warranted on experienced sonographers using CPD for IO catheter confirmation.