Related Experiment Video
Updated: Sep 25, 2025

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Three-dimensional morphometric analysis of the lateral clavicle and acromion: Implications for surgical treatment
Martin Zenker1, Javad Shamsollahi1, André Galm1
1R&D Department, DePuy Synthes, Zuchwil, Switzerland.
Objectives:
Dislocations and periarticular fractures of the acromioclavicular joint are common injuries of the shoulder girdle. When surgical intervention is indicated, subacromial support is one option to restore the alignment between scapula and the distal/lateral clavicle. Devices used for subacromial support rely on a form of subacromial 'hook'. The shape, inclination and orientation of which is often mismatched to the anatomy of the inferior surface of the acromion, which may lead to painful acromial osteolysis and rotator cuff abrasion causing impingement. The primary goal of this study was to characterize the geometrical parameters of the acromion and distal clavicle, and their orientation at the acromioclavicular joint.
Methods:
Computed tomography scans of 120 shoulders were converted into digital three-dimensional models. Measurements of the acromion inclination and acromion width relative to the torsional angle as well as the clavicle depth were taken. A numerical optimization of the anatomical parameters (including torsional and inclination angles, height and width) was performed to find the combination of those parameters with the lowest interpatient variability.
Results:
The mean clavicle depth was found to be 11.1 mm. The mean acromion width was 27 mm. The combination of torsional and inclination angles with lowest interpatient variability was found at 80° and 16°, respectively.
Conclusion:
There is a high interpatient variability in the morphology of the inferior surface of the acromion. Subacromial support using a 'hook' can be optimized for contact surface area, which should lead to fewer complications after the restoration of acromioclavicular orientation using acromial support strategies.