How to estimate R50% for cranial SRS/SRT cases with overlapping 50% isodose volumes: A proposed system
Dharmin D Desai1, Ivan L Cordrey2
1Department of Radiation Oncology, CHI Memorial Hospital, Chattanooga, Tennessee, USA.
A new method, the fair value estimate (FVE), systematically apportions overlapping 50% isodose clouds (IDC50%) in cranial stereotactic treatments. This approach ensures consistent assessment of intermediate dose spill for multiple targets, aiding adherence to proposed standards.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiosurgery
Background:
- Clinical stereotactic treatments with multiple targets often result in overlapping 50% isodose clouds (IDC50%).
- Accurately assessing dose conformity and intermediate dose spill becomes challenging in these overlapping scenarios.
- Existing methods struggle to consistently apportion dose when IDC50% from multiple planning target volumes (PTVs) intersect.
Purpose of the Study:
- To propose and validate a systematic method for apportioning overlapping IDC50% in single-isocenter cranial stereotactic treatments.
- To enable consistent evaluation of intermediate dose spill against emerging universal standards.
- To provide a reliable approach for cases involving multiple PTVs with intersecting IDC50%.
Main Methods:
- A novel "fair value estimate" (FVE) system was developed, apportioning IDC50% based on the relative surface area of each target volume.
- The FVE system's reasonableness was tested by comparing its results to actual R50% values in widely spaced target cases.
- The application of the FVE system to overlapping IDC50% scenarios was demonstrated.
Main Results:
- The FVE system demonstrated reasonable apportionment, yielding results similar to actual R50% in non-overlapping cases.
- The study provides a clear methodology for applying FVE to complex cases with overlapping IDC50%.
- The proposed method offers a consistent way to handle intermediate dose spill apportionment.
Conclusions:
- The fair value estimate (FVE) provides a systematic and consistent method for apportioning overlapping IDC50% in multi-target cranial stereotactic treatments.
- This method facilitates accurate assessment of intermediate dose spill, crucial for meeting proposed universal standards.
- The FVE system is recommended for consideration by the stereotactic community to address challenges in dose apportionment.