Related Experiment Video
Updated: Sep 24, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Impact of thyroid gland shielding on radiation doses in dental cone beam computed tomography with small and medium
Melanie Grüning1, Juha Koivisto2, James Mah3
1Charite Universitatsmedizin Berlin.
Objective:
The purpose of this study was to evaluate the impact of thyroid gland shielding on radiation doses in dental cone beam computed tomography (CBCT) with small and medium fields of view (FOVs).
Study Design:
Six CBCT protocols were investigated by exposing an adult anthropomorphic male phantom head without and with thyroid shielding, using 4 small (4 × 5 cm) and 2 medium (10 × 6 cm) FOVs. Twenty metal oxide semiconductor field-effect transistor dosimeters were placed in the phantom head to measure absorbed doses and calculate equivalent doses at 11 sites. Effective doses were calculated based on the tissue weighting factors in International Commission on Radiological Protection Publication 103. The data were analyzed using the independent samples t test.
Results:
Thyroid gland shielding led to significant equivalent dose reductions in many tissues for all protocols. Equivalent dose reductions to the thyroid were significant in all 6 protocols (P ≤ .037). Significant reduction depended on the FOV and ranged between 24.5% and 42.6% for the thyroid gland and 4.9% and 34.5% for other tissues and organs. Effective doses were significantly lower in all protocols (P ≤ .016).
Conclusions:
Thyroid gland shielding protects the thyroid gland and other organs and should be utilized with all CBCT examinations where feasible.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation
X-ray Imaging