Related Experiment Video
Updated: Sep 23, 2025

A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Intraoperative Quantification of MDS-UPDRS Tremor Measurements Using 3D Accelerometry: A Pilot Study
Annemarie Smid1, Jan Willem J Elting2,3, J Marc C van Dijk1
1Department of Neurosurgery, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands.
Objective accelerometry can measure Parkinson's disease (PD) tremor, translating the Movement Disorder Society—Unified PD Rating Scale (MDS-UPDRS) into an objective method. While showing concordance, accelerometric kinetic tremor measures had discrepancies, highlighting potential for reduced rater dependency.
Area of Science:
- Neurology
- Biomedical Engineering
- Movement Disorders
Background:
- The Movement Disorder Society—Unified PD Rating Scale (MDS-UPDRS) is standard for Parkinson's disease (PD) tremor assessment.
- MDS-UPDRS suffers from subjectivity and rater dependency.
- Objective motor measurements via accelerometry offer a potential solution.
Purpose of the Study:
- To translate MDS-UPDRS tremor tests into an objective scoring method using 3D accelerometry.
- To develop and validate an algorithm for measuring and classifying PD tremor based on MDS-UPDRS criteria.
Main Methods:
- 28 PD patients and 26 healthy controls underwent MDS-UPDRS tremor rating and 3D accelerometry of index fingers.
- Measurements were taken in an off-medication state.
- Quantitative metrics like tremor amplitude and power spectral density were calculated.
Main Results:
- A substantial concordance (>69.6%) was found between accelerometric and MDS-UPDRS tremor scores.
- Trends aligned with previous studies on tremor amplitude and MDS-UPDRS scores.
- Accelerometric kinetic tremor measures showed poor association (R2 < 0.3) with MDS-UPDRS, attributed to noise in healthy controls.
Conclusions:
- MDS-UPDRS tremor tests can be objectively translated using accelerometry.
- Discrepancies exist between objective kinetic tremor measures and subjective MDS-UPDRS ratings.
- This technology promises reduced rater dependency and objective intraoperative tremor monitoring.
More Related Videos
04:06Author Spotlight: Enhancing Remote Rehabilitation with Virtual Reality and Electromyography
Published on: January 12, 2024
11:06A Human-machine-interface Integrating Low-cost Sensors with a Neuromuscular Electrical Stimulation System for Post-stroke Balance Rehabilitation
Published on: April 12, 2016