An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Bone Cements Used for Hip Prosthesis Fixation: The Influence of the Handling Procedures on Functional Properties
Alina Robu1, Robert Ciocoiu1, Aurora Antoniac1
1Faculty of Material Science and Engineering, University Politehnica of Bucharest, 313 Splaiul Independentei, District 6, 060042 Bucharest, Romania.
Abstract:
The failure of hip prostheses is a problem that requires further investigation and analysis. Although total hip replacement is an extremely successful operation, the number of revision surgeries needed after this procedure is expected to continue to increase due to issues with both bone cement types and cementation techniques (depending on the producer). To conduct a comparative analysis, as a surgeon prepared the bone cement and introduced it in the body, this study's team of researchers prepared three types of commercial bone cements with the samples mixed and placed them in specimens, following the timeline of the surgery. In order to evaluate the factors that influenced the chemical composition and structure of each bone cement sample under specific intraoperative conditions, analyses of the handling properties, mechanical properties, structure, and composition were carried out. The results show that poor handling can impede prosthesis-cement interface efficacy over time. Therefore, it is recommended that manual mixing be avoided as much as possible, as the manual preparation of the cement can sometimes lead to structural unevenness.