Related Experiment Video
Updated: Sep 22, 2025

Implementation of a Real-Time Psychosis Risk Detection and Alerting System Based on Electronic Health Records using CogStack
Published on: May 15, 2020
Failure Mode and Effects Analysis (FMEA) at the preanalytical phase for POCT blood gas analysis: proposal for a
Viviane Van Hoof1, Suzanne Bench2, Antonio Buño Soto3
1Faculty of Medicine and Health Sciences, University of Antwerp, Wilrijk, Belgium.
Objectives:
Proposal of a risk analysis model to diminish negative impact on patient care by preanalytical errors in blood gas analysis (BGA).
Methods:
Here we designed a Failure Mode and Effects Analysis (FMEA) risk assessment template for BGA, based on literature references and expertise of an international team of laboratory and clinical health care professionals.
Results:
The FMEA identifies pre-analytical process steps, errors that may occur whilst performing BGA (potential failure mode), possible consequences (potential failure effect) and preventive/corrective actions (current controls). Probability of failure occurrence (OCC), severity of failure (SEV) and probability of failure detection (DET) are scored per potential failure mode. OCC and DET depend on test setting and patient population e.g., they differ in primary community health centres as compared to secondary community hospitals and third line university or specialized hospitals. OCC and DET also differ between stand-alone and networked instruments, manual and automated patient identification, and whether results are automatically transmitted to the patient's electronic health record. The risk priority number (RPN = SEV × OCC × DET) can be applied to determine the sequence in which risks are addressed. RPN can be recalculated after implementing changes to decrease OCC and/or increase DET. Key performance indicators are also proposed to evaluate changes.
Conclusions:
This FMEA model will help health care professionals manage and minimize the risk of preanalytical errors in BGA.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Steps in Outbreak Investigation
Pre-Procedural Guidelines for Assessing Blood Pressure
Methods of Documentation VI: Case Management Model
For example, a patient with a chronic...