Comparison of two new integrated SS-OCT tomography and biometry devices
Nir Sorkin1, Asaf Achiron, Muhammad Abumanhal
1From the Department of Ophthalmology, Tel Aviv Medical Center and Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel (Sorkin, Achiron, Abumanhal, Cohen, Gutfreund, Mandelblum, Varssano, Levinger); Department of Ophthalmology, Shaare Zedek Medical Center and the Hebrew University-Hadassah Medical School, Jerusalem, Israel (Abulafia).
Purpose:
To evaluate the agreement in biometry measurements and intraocular lens (IOL) power calculations between 2 integrated swept-source optical coherence tomography (SS-OCT) tomography/biometry devices: the Eyestar 900 and the Anterion.
Setting:
Tel Aviv Medical Center, Tel Aviv, Israel.
Design:
Retrospective comparative study.
Methods:
113 eyes of 66 consecutive patients who were evaluated prior to cataract surgery using both devices on the same visit were included. Measurements of axial length, anterior and posterior keratometry, anterior chamber depth, corneal diameter (CD), central corneal thickness, and lens thickness were recorded from both devices and analyzed for agreement and correlation. IOL power calculations using Barrett Universal II formula were compared between the devices in addition to prediction error (PE) and the mean absolute error (MAE).
Results:
The mean age was 71.6 ± 9.8 years. Clinically, the mean differences in measured variables were small. Measurement correlation and agreement between the devices were excellent for all variables, except for CD (Spearman r = 0.763, P < .001, and Bland-Altman 95% limits of agreement -0.21 to +0.65 mm). Good agreement was found between the devices in IOL power calculations, PE, and MAE, and the agreement was better when using the Eyestar 900 reflective keratometry measurements than the Eyestar 900 SS-OCT keratometry measurements (IOL power agreement within ±0.5 diopter (D) in 90.5% [n = 95] and 81.0% [n = 85], respectively [ P = .076], and within ±0.25 D in 64.8% [n = 68] and 44.8% [n = 47], respectively [ P = .005]).
Conclusions:
The 2 integrated SS-OCT tomography/biometry devices had good agreement in biometry measurements and IOL power calculations. Agreement was better when the Eyestar 900 reflective anterior keratometry measurements were used rather than the Eyestar 900 SS-OCT anterior keratometry measurements. Measurements of CD had reduced agreement and were not interchangeable.
More Related Videos
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
09:56Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography