Two-step aberration correction: application to transcranial histotripsy
Ning Lu1, Timothy L Hall1, Jonathan R Sukovich1
1Department of Biomedical Engineering, University of Michigan, Ann Arbor, United States of America.
Physics in Medicine and Biology
|May 24, 2022
Summary
This study enhances aberration correction for transcranial histotripsy by combining CT-based and acoustic cavitation emission methods. The improved 2-step approach significantly refines focal accuracy for brain therapy.
Area of Science:
- Medical Physics
- Acoustic Imaging
- Neurosurgery
Background:
- Transcranial histotripsy requires accurate focusing through the skull.
- Skull bone heterogeneity causes significant focal distortion.
- Existing aberration correction methods have limitations in clinical settings.
Purpose of the Study:
- To improve a 2-step aberration correction (AC) method for transcranial histotripsy.
- To develop a clinically applicable AC workflow.
- To compensate for focal distortion caused by skull bone heterogeneity.
Main Methods:
- Implemented a 2-step AC workflow: computed tomography (CT)-based analytical approach followed by acoustic cavitation emission (ACE)-based approach.
- Utilized a 700 kHz, 360-element hemispherical transducer array for cavitation generation and ACE shockwave acquisition.
- Investigated forward and backward ray-tracing models for CT-based AC and cross-correlation for ACE-based AC.
Main Results:
- The 2-step AC achieved 90 ± 7% peak focal pressure compared to hydrophone correction.
- Reduced average focal shift from 0.84 mm to 0.30 mm.
- Decreased average focal volume from 10.6 mm³ to 2.0 mm³ compared to no AC.
Conclusions:
- The 2-step AC method provides superior refocusing compared to CT-AC or ACE-AC alone.
- This improved method can be implemented in real-time for transcranial histotripsy.
- The workflow is suitable for clinical application in brain therapy.