Tunable laser spectroscopy for carbon dioxide capnography and water vapor sensing inside a breathing mask:
Lance E Christensen1, Kamjou Mansour1, Joachim D Pleil2
1Jet Propulsion Laboratory, California Institute of Technology, Pasadena, CA, United States of America.
Abstract:
Tunable laser spectroscopy (TLS) near 2683 nm was used to measure carbon dioxide and water vapor inside a pilot mask during jet fighter flights. Measurement frequency was 100 Hz in order to capture breathing profiles and other gas flow dynamics. Analysis of the full inhalation and exhalation breathing cycle allowed precise monitoring of breathing performance and interaction of the pilot with the life-support system. Measurements revealed dynamic phenomena pertaining to mechanical gas flow and pilot respiration that may be used to understand gas delivery stresses imposed upon the pilot and pilot physiology during flight. Typically, such measurements are made with non-dispersive infrared instrumentation for only carbon dioxide with intrinsic challenges regarding time and optical resolution. The TLS approach is a major advance because the sensor is placed directly into the mask improving its time response and enabling use of water vapor measurements that are less impacted from memory effects. This article presents the implementation of TLS and shows highly time-resolved pilot breathing data for high-performance aircraft tests.
More Related Videos
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Neural Control of Respiration
Respiratory Centers in the Brainstem
Two primary areas comprise the respiratory center: the medullary respiratory center in the medulla oblongata and the pontine respiratory group in the pons. The...