Electrowetting-based Digital Microfluidics Platform for Automated Enzyme-linked Immunosorbent Assay
Published on: February 23, 2020
4-Dimethylaminoantipyrine as a Broad Electrochemical Indicator for Immunosensors Platform
Francielli C C Melo1, Renata P Alves2, Anderson L Valle1
1Laboratory of Nanobiotechnology, Institute of Biotechnology, Federal University of Uberlândia, Avenida Amazonas s/n Bl. 2E, sl. 248, Uberlândia 38402-022, Minas Gerais, Brazil.
Abstract:
Here, we describe 4-dimethylaminoantipyrine (4-DMAA)-mediated interfacing as a broad biochemical indicator to stabilize and promote the higher response of electrodes for immunological detection. We hypothesized that the improved biological interactions of 4-DMAA with electrodes and biological samples may be due to the interaction properties of the benzene and pyrazole chemical groups with graphite and proteins, respectively. In order to demonstrate that 4-DMAA could be used as a general indicator in electrochemical immunoassays, we used peptides as probes for the diagnosis of four neglected tropical infectious diseases Tegumentary leishmaniasis, Visceral leishmaniasis, Strongyloidiasis, and Leprosy on commercial graphite screen-printed electrodes. 4-DMAA oxidation was used to indicate specific biological recognition between the epitope-based peptide and serum immunoglobulin G (IgG) from infected patients. We demonstrated that 4-DMAA should be incorporated into the electrodes prior to serum application, which avoids interference with its sensitivity and specificity. In addition, 4-DMAA oxidizes at a low anodic potential, and the oxidation peak is useful for detecting proteins in biological fluids. In summary, we have successfully demonstrated the broad application of 4-DMAA as a general indicator for the specific diagnosis of four infectious diseases in electrochemical immunosensors. Such a strategy is quite advantageous for indirect detection of proteins that lack electrochemical activities or are spatially inaccessible on the electrode surface. This new indicator opens a new avenue for monitoring biological recognition, especially for immunosensors.