High-Resolution Single Tooth MRI With an Inductively Coupled Intraoral Coil-Can MRI Compete With CBCT?
Tim Hilgenfeld1, Muhammad Abdullah Saleem1, Franz Sebastian Schwindling2
1From the Departments of Neuroradiology.
Objectives:
The aims of this study were to quantify T1/T2-relaxation times of the dental pulp, develop a realistic tooth model, and compare image quality between cone-beam computed tomography (CBCT) and high-resolution magnetic resonance imaging (MRI) of single teeth using a wireless inductively coupled intraoral coil.
Methods:
T1/T2-relaxometry was performed at 3 T in 10 healthy volunteers (283 teeth) to determine relaxation times of healthy dental pulp and develop a realistic tooth model using extracted human teeth. Eight MRI sequences (DESS, CISS, TrueFISP, FLASH, SPACE, TSE, MSVAT-SPACE, and UTE) were optimized for clinically applicable high-resolution imaging of the dental pulp. In model, image quality of all sequences was assessed quantitatively (contrast-to-noise ratio) and qualitatively (visibility of anatomical structures and extent of susceptibility artifacts using a 5-point scoring scale). Cone-beam computed tomography served as the reference modality for qualitative assessment. Statistical analysis was performed using 2-way analysis of variance, Fisher exact test, and Cohen κ.
Results:
In vivo, relaxometry of dental pulps revealed T1/T2 relaxation times at 3 T of 738 ± 100/171 ± 36 milliseconds. For all sequences, an isotropic resolution of (0.21 mm) 3 was achieved, with acquisition times ranging from 6:19 to 8:02 minutes. In model, the highest contrast-to-noise ratio values were observed for UTE, followed by TSE and CISS. The best image/artifact quality, however, was found for DESS (mean ± SD: 1.3 ± 0.3/2.2 ± 0.0), FLASH (1.5 ± 0.3/2.4 ± 0.1), and CISS (1.5 ± 0.4/2.5 ± 0.1), at a level comparable to CBCT (1.2 ± 0.3/2.1 ± 0.1).
Conclusions:
Optimized MRI protocols using an intraoral coil at 3 T can achieve an image quality comparable to reference modality CBCT within clinically applicable acquisition times. Overall, DESS revealed the best results, followed by FLASH and CISS.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging