Artificial Intelligence-Based Stethoscope for the Diagnosis of Aortic Stenosis
Tamer Ghanayim1, Lior Lupu2, Sivan Naveh2
1Department of Cardiology, Lady Davis Carmel Medical Center, Haifa, Israel.
Background:
The diagnostic accuracy of the stethoscope is limited and highly dependent on clinical expertise. Our purpose was to develop an electronic stethoscope, based on artificial intelligence (AI) and infrasound, for the diagnosis of aortic stenosis (AS).
Methods:
We used an electronic stethoscope (VoqX; Sanolla, Nesher, Israel) with subsonic capabilities and acoustic range of 3-2000 Hz. The study had 2 stages. In the first stage, using the VoqX, we recorded heart sounds from 100 patients referred for echocardiography (derivation group), 50 with moderate or severe AS and 50 without valvular disease. An AI-based supervised learning model was applied to the auscultation data from the first 100 patients used for training, to construct a diagnostic algorithm that was then tested on a validation group (50 other patients, 25 with AS and 25 without AS). In the second stage, conducted at a different medical center, we tested the device on 106 additional patients referred for echocardiography, which included patients with other valvular diseases.
Results:
Using data collected at the aortic and pulmonic auscultation points from the derivation group, the AI-based algorithm identified moderate or severe AS with 86% sensitivity and 100% specificity. When applied to the validation group, the sensitivity was 84% and specificity 92%; and in the additional testing group, 90% and 84%, respectively. The sensitivity was 55% for mild, 76% for moderate, and 93% for severe AS.
Conclusion:
Our initial findings show that an AI-based stethoscope with infrasound capabilities can accurately diagnose AS. AI-based electronic auscultation is a promising new tool for automatic screening and diagnosis of valvular heart disease.
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...