Published on: February 19, 2020
Ultrafast Measurement of Metformin in the Clinical Setting Using Probe Electrospray Ionization Mass Spectrometry
Pauline Griffeuille1, Souleiman El Balkhi1, Sandra Bodeau2
1Department of Pharmacology and Toxicology, Limoges University Hospital, 2 Avenue Martin Luther King, Limoges 87042, France.
Abstract:
Metformin (MtF) is a treatment used for type 2 diabetes. Lactic acidosis (LA) is a frequent complication that can be either induced by or associated with elevated MtF plasma concentrations. When coupled with a mass spectrometry (MS) system, the probe electrospray ionization (PESI) method allows direct and rapid analysis of different types of matrices without pretreatment. In this study, we developed a PESI-MS method for the determination of MtF in plasma. We used a tandem mass spectrometer equipped with a PESI source in the reaction monitoring mode for the quantitation of MtF. MtF-d6 was chosen as the internal standard (IS), following an isotope dilution (ID) approach. The method was fully validated with six concentration levels (0.5-50 mg/L). The matrix effect was evaluated for each level, and the specificity was tested with a mix of potential co-medications. Using patient samples, the performance was compared with two classical LC-MS-MS and LC-diode array detector (DAD) methods used in external labs. Sample preparation consisted in mixing 10 µL plasma in 1,000 µL ethanol/ammonium formate buffer including MtF-d6 at a fixed concentration of 5 mg/L. The total run time was 0.31 min. ID gave satisfactory results of accuracy and precision (min-max: -12.1 to 15.8% and 1.0-17.1%, respectively). The matrix effect was fully corrected by the internal standard (bias < 1%). The specificity study also reported satisfactory results. Finally, in a representative group of 29 patients (55% with a concentration <5 mg/L, 38% with a concentration >5 mg/L and 7% not detected), we observed almost identical results when comparing LC-DAD and LC-MS-MS to PESI-MS (r2 > 0.99). We propose a specific, sensitive, accurate and ultrafast solution for the measurement of MtF in patient plasma, with no sample preparation or calibration curve building. This could be helpful in a core lab when rapid diagnosis of LA is needed.
More Related Videos
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
MALDI-TOF Mass Spectrometry
Matrix-assisted laser desorption ionization (MALDI) is a commonly...