Related Experiment Video
Updated: Sep 21, 2025

Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
Endoscopic optical coherence tomography angiography using inverse SNR-amplitude decorrelation features and
Lin Yao1, Huakun Li1, Kaiyuan Liu1
1State Key Laboratory of Modern Optical Instrumentation, College of Optical Science and Engineering, Zhejiang University, Hangzhou, China.
Background:
Angiogenesis is closely associated with tumor development and progression. Endoscopic optical coherence tomography angiography (OCTA) enables rapid inspection of mucosal 3D vasculature of inner organs in the early-stage tumor diagnosis; however, it is limited by instabilities of the optical signal and beam scanning.
Methods:
In the phase-unstable swept source OCTA (SS-OCTA), amplitude decorrelation was used to compute the motion-induced changes as motion contrast. The influence of the random noise-induced amplitude fluctuations on decorrelation was characterized as a function of inverse signal-to-noise ratio (SNR) with a multi-variate time series (MVTS) model and statistical analysis. Then, the noise-induced decorrelation artifacts in static tissue regions were eliminated by applying a flow mask based on the statistical relation between inverse SNR (iSNR) and amplitude decorrelation (IDa), which was named IDa-OCTA. In addition, a distal stepwise raster scan was realized with a low-voltage electrothermal micro-electro-mechanical system (ET-MEMS)-based catheter for endoscopic imaging, whereby the stable and repeatable B-scans at each step suppressed the decorrelation noise induced by the spatial mismatch between paired scans.
Results:
The derived IDa relation was validated through numerical simulation and flow phantom experiments. In vivo human buccal mucosa imaging was performed to demonstrate the endoscopic IDa-OCTA imaging. In this, the subsurface structure and vasculature were visualized in a rapid and depth-resolved manner.
Conclusions:
The rapid 3D vasculature visualization realized by the endoscopic IDa-OCTA improves the diagnosis of early tumors in internal organs.