Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
3.8K
Evaluation of septal perforation size using a novel measuring device
Stefan Hamilton1, Brian W Rotenberg1, Leigh J Sowerby2
1Department of Otolaryngology-Head and Neck Surgery, St. Joseph's Healthcare Centre, Western University, 268 Grosvenor St., London, ON, N6A 4V2, Canada.
Summary
New 3D-printed sizers significantly improve the accuracy of nasal septal perforation measurements compared to traditional methods. These novel instruments enhance ease of use and offer potential for improved patient care.
Area of Science:
- Otolaryngology
- Medical Device Development
- Biomedical Engineering
Background:
- Nasal septal perforations impact airflow and require management, often with septal button prostheses.
- Accurate measurement of perforations is crucial for effective prosthesis fitting and patient comfort.
- Current clinical methods for measuring septal perforations can be challenging and inaccurate.
Purpose of the Study:
- To evaluate the accuracy and ease of use of two novel 3D-printed instruments for measuring nasal septal perforations.
- To compare the performance of these new sizers against traditional measurement techniques.
Main Methods:
- Two types of 3D-printed sizers (serial and graded) were developed.
- Nasal septal perforations were created in cadaver heads.
- Otolaryngology professionals measured perforations using eyeball estimation, a ruler, and the novel sizers.
- Accuracy was defined as ±1 mm of the true measurement; statistical analyses included Chi-square and ANOVA.
Main Results:
- The novel sizers demonstrated significantly higher accuracy in measuring perforation length and height compared to eyeball estimation and ruler methods.
- The eyeball method significantly overestimated perforation length and was less accurate than the other methods.
- No correlation was found between participant experience and measurement accuracy.
- Participants showed a clear preference for the novel sizers due to ease of use.
Conclusions:
- The developed 3D-printed sizers are more accurate and user-friendly than traditional methods for measuring nasal septal perforations.
- Wider adoption of these sizers could lead to improved patient outcomes and satisfaction.
- Further study and implementation of these devices are recommended for advancing septal perforation management.