Related Experiment Video
Updated: Sep 20, 2025

Interventional Diagnostic Procedure: A Practical Guide for the Assessment of Coronary Vascular Function
Published on: March 15, 2022
The DERIVO 2heal embolisation device: A technical report using single antiplatelet therapy for intracranial
Lukas Goertz1, Michael Schoenfeld1, David Zopfs1
1Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
Abstract:
The novel DERIVO 2heal Embolisation Device (Acandis, Pforzheim, Germany) is a flexible, fully radiopaque flow-diverter with a fibrin-based nano-coating, which is supposed to make the device inert to the coagulation cascade. We report a case of pseudoaneurysm treatment with this device under single anti-platelet therapy (SAPT). A female patient underwent endoscopic surgery for chronic rhinosinusitis. During surgery, the lateral wall of the sphenoid wall and the adjacent internal carotid artery was injured, leading to massive hemorrhage, which was treated with compression and tamponades. Eleven days after the procedure, the patient developed a rapidly growing pseudoaneurysm at the injury site with a relevant risk of upcoming aneurysm rupture. The aneurysm was treated by implantation of a DERIVO 2heal. A SAPT regimen was chosen due to the recent massive hemorrhage and anticipating enhanced progressive aneurysm occlusion. There were no thromboembolic complications and there was no morbidity. After 5 days, the pseudoaneurysm was completely occluded. The observations of the presented case are promising and warrant further investigation of this device.