Monitoring of Heated Wire Humidifier Hygrometric Performances With Heater Plate Temperature
Related Experiment Video
Updated: Sep 20, 2025

Fabrication of Microscope Stage for Vertical Observation with Temperature Control Function
Published on: July 31, 2019
François Lellouche1, Serge Simard2, Pierre-Alexandre Bouchard2
1Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Québec, Québec, Canada. francois.lellouche@criucpq.ulaval.ca.
Background:
Under-humidification and associated complications may occur with heated humidifiers. Hygrometric performances of heated wire humidifiers are reduced by high ambient and high outlet ventilator temperatures. Currently, there is no reliable monitoring tool to evaluate humidification performances of heated wire humidifiers in the daily practice. We sought to demonstrate the relation between humidity delivered by heated wire humidifiers and different parameters that could be used to monitor humidity of gas delivered to subjects.
Methods:
On a bench test, we measured heater plate temperature, inlet chamber temperature, and delivered humidity with MR850 system. Temperature displayed on the humidifier was also recorded. The measurements were performed at different ambient temperatures and five minute ventilation levels (5, 7.5, 10, 12.5, 15 L/min). Inlet chamber temperatures varied from 20-40°C. In each condition, hygrometric measurements with the psychrometric method were performed at steady state.
Results:
We performed 279 measurements at steady state including all conditions. We found a good correlation between heater plate temperature and absolute humidity delivered (R2 = 0.82). This relationship was hardly affected by ambient temperature, but minute ventilation had more effect. For different minute ventilations, the correlation between heater plate temperature and absolute humidity delivered was very good with coefficient of determination R2 from 0.87-0.98. Heater plate temperature > 62°C was a good predictor of absolute humidity delivered > 30 mg H2O/L (area under the curve = 0.96, sensibility 79%, specificity 94%). No correlation existed between humidity delivered and the outlet chamber temperature (displayed on the humidifier).
Conclusions:
In this bench study, we have shown a good correlation between heater plate temperature and humidity delivered with a heated wire humidifier. This means that a "hidden" hygrometer is built into the heated wire humidifier. Heater plate temperature should be used as a surrogate of humidity to improve the humidification monitoring.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...