Related Experiment Video
Updated: Sep 20, 2025

Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
Alternative instrument for the evaluation of handgrip strength in Duchenne muscular dystrophy
Mariana Angélica de Souza1, Edson Zangiacomi Martinez2, Elisângela Aparecida da Silva Lizzi3
1Department of Health Science, Ribeirao Preto Medical School, University of São Paulo, Av. Bandeirantes, 3900, Campus, Ribeirão Preto, SP, 14049-900, Brazil.
Background:
The commonly used dynamometers can be ineffective in evaluating handgrip in patients with Duchenne muscular dystrophy (DMD), especially children with generalized muscle weakness. The aim of this study was to analyze whether the modified sphygmomanometer is an effective instrument for handgrip strength evaluation in patients with DMD, during different stages of the disease.
Method:
The handgrip strength of 33 patients was evaluated by the Jamar dynamometer and the modified sphygmomanometer. Motor function was evaluated by the Motor Function Measurement (MFM) scale. Four evaluations, with a six-month interval between each, were performed: Evaluation 1 (N = 33), Evaluation 2 (N = 24), Evaluation 3 (N = 15), and Evaluation 4 (N = 8). A linear regression model with mixed effects was used for the longitudinal data and descriptive analysis of strength for all four evaluations.
Result:
The first evaluation data presented very high correlations between the dynamometer and the modified sphygmomanometer (r = 0.977; p < 0.001). The longitudinal analysis showed a significant difference between Evaluation 1 and the other handgrip strength evaluations obtained using the dynamometer (p < 0.05) but not the modified sphygmomanometer (p > 0.05). Null values were obtained only when using the dynamometer device.
Conclusion:
The modified sphygmomanometer seems to be more suitable than the dynamometer for measuring handgrip strength in all stages of DMD.