Measurement of Pulse Propagation Velocity, Distensibility and Strain in an Abdominal Aortic Aneurysm Mouse Model
Published on: February 23, 2020
Ultrasound Transducer Pressure: An Unexplored Source of Abdominal Aortic Aneurysm Measurement Error
Qasam Mohammed Ghulam1, Morten Bo Søndergaard Svendsen2, Alexander Hakon Zielinski3
1Department of Vascular Surgery, Rigshospitalet, Copenhagen, Denmark.
Abstract:
This study was aimed at quantifying the abdominal aortic aneurysm (AAA) compression phenomenon and assessing the use of a new tool to estimate transducer pressure in vivo. In this cross-sectional study, 47 participants with AAA and a median anterior-to-posterior (AP) AAA diameter of 46 mm (range: 30-76 mm) were included. The majority of the patients were overweight with body mass indexes >25 (33/47, 70%). A standardized ultrasound (US)-compatible gel pad, with a pre-defined thickness (15 mm) and mass (150 g), was interposed between the US transducer and participant to estimate the applied transducer pressure. A firm transducer pressure significantly lowered the median AP diameter from 46.1 mm (range: 29.7-76.3) to 39.4 mm (range: 21.7-67.5) (p = 0.001). The mean diameter bias between light and firm transducer pressure was 6.1 mm (95% confidence interval: 4.9-7.3). The applied transducer pressure varied significantly and could be measured by translating the deformation of a low-tech gel pad interposed between the US transducer and the participant.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Aneurysm II: Clinical Manifestations and Diagnostic Studies