Comparative Analysis of Physical Activity Detected via an External Accelerometer and Cardiac Implantable Electronic
Chun-Kai Chen1, Li-Ying Cheng1,2, Shan-Wei Hsu1
1Division of Cardiology, Department of Internal Medicine, National Taiwan University Hospital Hsin-Chu Branch, Hsinchu, Taiwan.
Background:
Physical activity (PA) has become an important health issue for decades. Cardiovascular implantable electronic devices (CIEDs) have built-in PA-recording functions. We aimed to compare PA measurements using an external accelerometer (ActiGraph GT3X+) and internal accelerometers (Abbott, Biotronik, and Medtronic CIEDs).
Methods:
This was a prospective, single-center observational study. The device-measured 7-day average PA was collected, and GT3X+ -measured 7-day average PA was used as the gold-standard, including all daily observations of activity. Pearson's correlation coefficients were used to compare the correlations between GT3X+ -measured and CIED-measured PA. Bland-Altman plots were used to analyze measurement agreement, and intraclass correlation coefficients were used to analyze reliability.
Results:
In total, 720 patients treated with CIEDs were surveyed between November 2020 and April 2021, 60 of them were analyzed after patient screening by our protocol. Each manufacturer included 20 patients for the final analysis. The CIED-measured PAs of Abbott, Biotronik, and Medtronic were 3.0 ± 1.5, 2.6 ± 1.8, and 3.8 ± 2.5 h per day, respectively; the GT3X+ -measured PAs were 6.9 ± 2.8, 6.0 ± 2.4, and 6.4 ± 2.5 h per day, respectively. Moderate and significant correlations were found in patients using Abbott, Biotronik, and Medtronic CIEDs (r = 0.534, p = 0.015; r = 0.465, p = 0.039; r = 0.677, p = 0.001, respectively). Bland-Altman plots and intraclass correlation coefficients both showed a significant correlation and reliability between the average PA measured by GT3X+ and CIEDs (hours per day).
Conclusion:
Although the PA recording function of CIEDs includes a single-axis accelerometer, it has a moderate correlation compared with the triaxial accelerometer of the GT3X+. However, CIEDs seem to underestimate PA for 3-4 h compared to the GT3X+.
More Related Videos
05:59Visualization of Intensity Levels to Reduce the Gap Between Self-Reported and Directly Measured Physical Activity
Published on: March 7, 2019
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring