Comparisons of Fifth-, Sixth-, and Seventh-Generation Continuous Glucose Monitoring Systems
John B Welsh1, Simon Psavko1, Xiaohe Zhang1
1Dexcom, Inc., San Diego, CA, USA.
Background:
Between-system differences for continuous glucose monitoring (CGM) devices have important clinical consequences.
Purpose:
Here we review attributes of Dexcom's fifth-, sixth-, and seventh-generation (G5, G6, and G7) CGM systems.
Methods:
Accuracy metrics were derived from preapproval trials of the three systems and compared after propensity score adjustments were used to balance baseline demographic characteristics. Metrics included mean absolute relative differences (MARD) between CGM and YSI values and the proportion of CGM values within 20% or 20 mg/dL of the YSI values ("%20/20"). Ease-of-use was evaluated by formal task analysis.
Conclusions:
Adjusted MARD and %20/20 agreement rates were 9.0%/93.1% (abdomen-placed G5), 9.9%/92.3% (abdomen-placed G6), 9.1%/93.2% (abdomen-placed G7), and 8.2%/95.3% (arm-placed G7). Task analysis favored G7 over earlier systems. Favorable clinical outcomes such as hemoglobin A1c reduction and hypoglycemia avoidance seen with G5 and G6 are anticipated with G7 use.
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Hypoglycemia and Glucagon
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...