Improvement in the Reliability of AOAC Official MethodSM 2012.15 for Iodine, Part-2
Naoto Hieda1, Motoyuki Nishida1, Ichirou Matsuno1
1Megmilk Snow Brand Co. Ltd, Quality Assurance Department, Central Food Analysis Laboratory, 1-1-2 Minamidai, Kawagoe, Saitama 350-1165, Japan.
Background:
In our previous study, the AOAC Official MethodSM2012.15 (Method 2012.15) was modified by adding carbon (as methanol) to the standard and final test solutions for accurate quantification of iodine in infant and special formulas after incomplete matrix oven digestion. However, when an inductively coupled plasma-mass spectrometry (ICP-MS) instrument from a different manufacturer was used, inaccurately low iodine content values were obtained using both the original and modified methods.
Objective:
To obtain more reliable values for infant formulas using varying ICP-MS instruments.
Methods:
The protocol was modified by changing the ratio of the pump tubing inner diameters (IDs) for the ICP-MS system or by optimizing the internal standards for Method 2012.15.
Results:
No significant improvement was observed by changing the ratio of the pump tubing IDs. In the optimization study of the internal standard, when germanium is used as the internal standard, the measurement result of NIST SRM® 1849 was within the certified value range. Furthermore, with the addition of methanol equivalent to 5% to the standard and final test solution based on our previous study, the recovery of the samples was within the certified range of SMPR 2012.008.
Conclusions:
This demonstrates that changing the internal standard to germanium and adding methanol equivalent to 5% to the standard and final test solution, can result in stable measurement even when the ICP-MS specifications were changed.
Highlights:
The modified Method 2012.15 provides reliable results even if the instrument is changed by changing the internal standard to germanium.
More Related Videos
08:21Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
09:54Chemoselective Preparation of 1-Iodoalkynes, 1,2-Diiodoalkenes, and 1,1,2-Triiodoalkenes Based on the Oxidative Iodination of Terminal Alkynes
Published on: September 12, 2018
Related Concept Videos
Redox Titration: Iodimetry and Iodometry
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Redox Titration: Overview
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Redox Titration: Other Oxidizing and Reducing Agents