High-resolution in vivo monophasic gastric slow waves to quantify activation and recovery profiles
Henry Han1, Leo K Cheng1,2, Niranchan Paskaranandavadivel1
1Auckland Bioengineering Institute, The University of Auckland, New Zealand.
Background:
Gastric bio-electrical slow waves are, in part, responsible for coordinating motility. Spatial dynamics about the recovery phase of slow wave recordings have not been thoroughly investigated due to the lack of suitable experimental techniques.
Methods:
A high-resolution multi-channel suction electrode array was developed and applied in pigs to acquire monophasic gastric slow waves. Signal characteristics were verified against biphasic slow waves recorded by conventional surface contact electrode arrays. Monophasic slow wave events were categorized into two groups based on their morphological characteristics, after which their amplitudes, activation to recovery intervals, and gradients were quantified and compared. Coverage of activation and recovery maps for both electrode types were calculated and compared.
Key Results:
Monophasic slow waves had a more pronounced recovery phase with a higher gradient than biphasic slow waves (0.5 ± 0.1 vs. 0.3 ± 0.1 mV·s-1 ). Between the 2 groups of monophasic slow waves, there was a significant difference in amplitude (1.8 ± 0.5 vs. 1.1 ± 0.2 mV), activation time gradient (0.8 ± 0.2 vs. 0.3 ± 0.1 mV·s-1 ), and recovery time gradient (0.5 ± 0.1 vs. 0.3 ± 0.1 mV·s-1 ). For the suction and conventional contact electrode arrays, the recovery maps had reduced coverage compared to the activation maps (4 ± 6% and 43 ± 11%, respectively).
Conclusions And Inferences:
A novel high-resolution multi-channel suction electrode array was developed and applied in vivo to record monophasic gastric slow waves. Slow wave recovery phase analysis could be performed more efficiently on monophasic signals compared with biphasic signals, due to the more identifiable recovery phases.