Breath-dependent pressure fluctuations in various constant- and variable-flow neonatal CPAP devices
Lorenz Auer-Hackenberg1, Patricia Stroicz1, Edda Hofstätter1
1Department of Neonatology, Division of Pediatrics and Adolescent Medicine, Paracelsus Medical Private University, Salzburg, Austria.
Objective:
In continuous positive airway pressure (CPAP) devices, pressure can be generated by two different mechanisms: either via an expiratory valve or by one or more jets. Valved CPAP devices are referred to as constant-flow devices, and jet devices are called variable-flow devices. Constant-flow CPAP devices are said to reduce the imposed work of breathing due to lower breath-dependent pressure fluctuations. The present study investigates the performance of various constant- and variable-flow CPAP devices in relation to breath-dependent pressure fluctuations.
Design:
Experimental study comparing the pressure fluctuations incurred by seven neonatal CPAP devices attached to an active neonatal lung model.
Methodology:
Spontaneous breathing was simulated using a tidal volume of 6 ml at pressure levels of 5, 7, and 9 mbar. The main outcomes were respiratory pressure fluctuations, tidal volume, and end-expiratory pressure.
Results:
All CPAP devices tested showed respiratory pressure fluctuations, varying from 0.631 to 3.466 mbar. The generated tidal volume correlated significantly with the pressure fluctuations (r = -0.947; p = 0.001) and varied between 5.550 and 6.316 ml. CPAP devices with jets showed no advantage over CPAP devices with expiratory valves. End-expiratory pressure in the nose deviated from the set pressure between -1.305 and 0.644 mbar and varied depending on whether the pressure was measured in the device or in the tube extending to the nose.
Conclusion:
During standard spontaneous breathing, breath-dependent pressure fluctuations in constant- and variable-flow devices are comparable. Pressure measurements taken in the tubing system can lead to a considerable deviation of the applied pressure.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Pulmonary Cycle: Exhalation
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Mechanical Ventilation I: Indication and Settings
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...