Related Experiment Video
Updated: Sep 6, 2025

A Multi-hole Cryovial Eliminates Freezing Artifacts when Muscle Tissues are Directly Immersed in Liquid Nitrogen
Published on: April 6, 2017
A cryogenic torsion balance using a liquid-cryogen free, ultra-low vibration cryostat
S M Fleischer1, M P Ross2, K Venkateswara2
1Department of Physics and Astronomy, Western Washington University, Bellingham, Washington 98225, USA.
Abstract:
We describe a liquid-cryogen free cryostat with ultra-low vibration levels, which allows for continuous operation of a torsion balance at cryogenic temperatures. The apparatus uses a commercially available two-stage pulse-tube cooler and passive vibration isolation. The torsion balance exhibits torque noise levels lower than room temperature thermal noise by a factor of about four in the frequency range of 3-10 mHz, limited by residual seismic motion and by radiative heating of the pendulum body. In addition to lowering thermal noise below room-temperature limits, the low-temperature environment enables novel torsion balance experiments. Currently, the maximum duration of a continuous measurement run is limited by accumulation of cryogenic surface contamination on the optical elements inside the cryostat.