A compact instrument for gamma-ray burst detection on a CubeSat platform II: Detailed design, assembly and validation
David Murphy1, Alexey Ulyanov1, Sheila McBreen1
1School of Physics & Centre for Space Research, University College Dublin, Dublin 4, Ireland.
Abstract:
The Gamma-ray Module, GMOD, is a miniaturised novel gamma-ray detector which will be the primary scientific payload on the Educational Irish Research Satellite (EIRSAT-1) 2U CubeSat mission. GMOD comprises a compact (25 mm 25 mm 40 mm) cerium bromide scintillator coupled to a tiled array of 4 4 silicon photomultipliers, with front-end readout provided by the IDE3380 SIPHRA. This paper presents the detailed GMOD design and the accommodation of the instrument within the restrictive CubeSat form factor. The electronic and mechanical interfaces are compatible with many off-the-shelf CubeSat systems and structures. The energy response of the GMOD engineering qualification model has been determined using radioactive sources, and an energy resolution of 5.4% at 662 keV has been measured. EIRSAT-1 will perform on-board processing of GMOD data. Trigger results, including light-curves and spectra, will be incorporated into the spacecraft beacon and transmitted continuously. Inexpensive hardware can be used to decode the beacon signal, making the data accessible to a wide community. GMOD will have scientific capability for the detection of gamma-ray bursts, in addition to the educational and technology demonstration goals of the EIRSAT-1 mission. The detailed design and measurements to date demonstrate the capability of GMOD in low Earth orbit, the scalability of the design for larger CubeSats and as an element of future large gamma-ray missions.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
06:46Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
Related Concept Videos
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...