Development of a new quantification method using partial volume effect correction for individual energy peaks in
Kosuke Yamashita1,2, Noriaki Miyaji1, Kazuki Motegi1
1Department of Nuclear Medicine, Cancer Institute Hospital of Japanese Foundation for Cancer Research, Tokyo, Japan.
Objectives:
Somatostatin receptor scintigraphy (SRS) using 111In-pentetreotide has no established quantification method. The purpose of this study was to develop a new quantitative method to correct the partial volume effect (PVE) for individual energy peaks in 111In-pentetreotide single-photon emission computed tomography (SPECT).
Methods:
Phantom experiments were performed to construct a new quantitative method. In the phantom experiments, a NEMA IEC body phantom was used. Acquisition was performed using two energy peaks (171 keV and 245 keV) on the SPECT/CT system. The volume of interest was set at each hot sphere and lung insert in the SPECT images of each energy peak, and the recovery coefficient (RC) was calculated to understand the PVE. A new quantitative index, the indium uptake index (IUI), was calculated using the RC to correct the PVE. The quantitative accuracy of the IUI in the hot sphere was confirmed. Case studies were performed to clarify the quantitative accuracy. In a case study, the relationship between the IUI and the Krenning score, which is used as a visual assessment, was evaluated for each lesion.
Results:
The obtained RCs showed that the energy peak at 171 keV was faster in recovering the effect of PVE than that at 245 keV. The IUI in the 17-mm-diameter hot sphere was overestimated by 4.8% and 8.3% at 171 keV and 245 keV, respectively, compared to the actual IUIs. The relationship between IUI and Krenning score was rs=0.773 (p<0.005) at sum, rs=0.739 (p<0.005) at 171 keV, and rs=0.773 (p<0.005) at 245 keV.
Conclusion:
We have developed a new quantification method for 111In-pentetreotide SPECT/CT using RC-based PVE correction for an individual energy peak of 171 keV. The quantitative accuracy of this method was high even for accumulations of less than 20 mm, and it showed a good relationship with the Krenning score; therefore, the clinical usefulness of IUI was demonstrated.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
06:31Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain
Published on: August 8, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET