Active-PET: a multifunctional PET scanner with dynamic gantry size featuring high-resolution and high-sensitivity
Amirhossein Sanaat1, Mahdi Jamalizadeh2, Hadi Khanmohammadi3
1Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, Geneva CH-1211, Switzerland.
Physics in Medicine and Biology
|July 8, 2022
Summary
The Active-PET scanner offers versatile imaging capabilities by adjusting detector configurations for optimal resolution and sensitivity. This multifunctional design overcomes the limitations of dedicated Positron Emission Tomography (PET) scanners, enhancing image quality for various applications.
Area of Science:
- Medical Imaging
- Nuclear Medicine
- Instrumentation
Background:
- Dedicated organ-specific PET scanners offer high resolution and sensitivity but are costly and space-intensive.
- Existing PET scanner designs lack flexibility for diverse clinical and research applications.
Purpose of the Study:
- To introduce and evaluate the Active-PET, a novel, multifunctional PET scanner concept.
- To demonstrate the capability of reconfigurable detector modules for variable field-of-view and performance optimization.
Main Methods:
- Monte Carlo simulations were employed to model a PET scanner with two sets of Lutetium Oxyorthosilicate (LSO(Ce)) detectors.
- Detector modules with varying pixel sizes and crystal thicknesses were utilized.
- Linear motors enabled radial displacement of detector modules to achieve variable gantry diameters.
Main Results:
- The simulated Active-PET achieved a maximum sensitivity of 15.98 kcps/MBq with a small gantry and high-sensitivity detectors.
- Optimal spatial resolution (2.2 mm tangential, 2.3 mm radial, 2.5 mm axial FWHM) was obtained with a small gantry and high-resolution detectors.
- A large-bore configuration demonstrated superior performance and image quality compared to the Biograph mCT in phantom studies.
Conclusions:
- The Active-PET concept presents a non-static PET scanner capable of switching between large/small fields-of-view and high-resolution/high-sensitivity imaging.
- This reconfigurable design offers a cost-effective and space-saving solution compared to multiple dedicated PET scanners.
- Active-PET holds potential for diverse applications including brain, breast, prostate, and image-guided therapy.