Automated Force-Coupled Ultrasound Method for Calibration-Free Carotid Artery Blood Pressure Estimation
Alex T Jaffe1, Rebecca E Zubajlo1, Luca Daniel1
1Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.
Abstract:
We develop, automate and evaluate a calibration-free technique to estimate human carotid artery blood pressure from force-coupled ultrasound images. After acquiring images and force, we use peak detection to align the raw force signal with an optical flow signal derived from the images. A trained convolutional neural network selects a seed point within the carotid in a single image. We then employ a region-growing algorithm to segment and track the carotid in subsequent images. A finite-element deformation model is fit to the observed segmentation and force via a two-stage iterative non-linear optimization. The first-stage optimization estimates carotid artery wall stiffness parameters along with systolic and diastolic carotid pressures. The second-stage optimization takes the output parameters from the first optimization and estimates the carotid blood pressure waveform. Diastolic and systolic measurements are compared with those of an oscillometric brachial blood pressure cuff. In 20 participants, average absolute diastolic and systolic errors are 6.2 and 5.6 mm Hg, respectively, and correlation coefficients are r = 0.7 and r = 0.8, respectively. Force-coupled ultrasound imaging represents an automated, standalone ultrasound-based technique for carotid blood pressure estimation, which motivates its further development and expansion of its applications.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.