Performance evaluation of a wrist-worn reflectance pulse oximeter during sleep
Hyunjun Jung1, Dongyeop Kim2, Wonkyu Lee1
1Samsung Electronics, Suwon, Republic of Korea.
Objectives:
To characterize and evaluate the estimation of oxygen saturation measured by a wrist-worn reflectance pulse oximeter during sleep.
Methods:
Ninety-seven adults with sleep disturbances were enrolled. Oxygen saturation was simultaneously measured using a reflectance pulse oximeter (Galaxy Watch 4 [GW4], Samsung, South Korea) and a transmittance pulse oximeter (polysomnography) as a reference. The performance of the device was evaluated using the root mean squared error (RMSE) and coverage rate. Additionally, GW4-derived oxygen desaturation index (ODI) was compared with the apnea-hypopnea index (AHI) derived from polysomnography.
Results:
The GW4 had an overall RMSE of 2.3% and negligible bias of -0.2%. A Bland-Altman density plot showed good agreement between the GW4 and the reference pulse oximeter. RMSEs were 1.65 ± 0.57%, 1.76 ± 0.65%, 1.93 ± 0.54%, and 2.93 ± 1.71% for normal (n = 18), mild (n = 21), moderate (n = 23), and severe obstructive sleep apnea (n = 35), respectively. The data rejection rate was 26.5%, which was caused by fluctuations in contact pressure and the discarding of data less than 70% of saturation. A GW4-ODI ≥5/h had the highest ability to predict AHI ≥15/h with sensitivity, specificity, accuracy, and area under the curve of 89.7%, 64.1%, 79.4%, and 0.908, respectively.
Conclusions:
This study evaluated the estimation of oxygen saturation by the GW4 during sleep. This device complies with both Food and Drug Administration and International Organization for Standardization standards. Further improvements in the algorithms of wearable devices are required to obtain more accurate and reliable information about oxygen saturation measurements.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation