Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation
Published on: September 2, 2025
Universal Method of Compatibility Assessment for Novel Ablation Technologies With Different 3D Navigation Systems
Luigi Pannone1, Ivan Eltsov1,2, Robbert Ramak1
1Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel - Vrije Universiteit Brussel, European Reference Networks Guard-Heart, Brussels, Belgium.
This study developed a bench-testing method to assess compatibility between ablation catheters and mapping systems. DiamondTemp ablation system (DTA) is compatible with Rhythmia electroanatomic mapping system (EAM) when using a specific configuration.
Area of Science:
- Medical Devices
- Electrophysiology
- Catheter Ablation
Background:
- Incompatibility between different manufacturers' ablation and mapping systems presents a clinical challenge.
- Devices are often used together despite lacking out-of-the-box cross-compatibility.
- A need exists for standardized compatibility testing methods.
Purpose of the Study:
- To develop a bench-testing method for assessing compatibility.
- To evaluate the compatibility between the DiamondTemp ablation system (DTA) and the Rhythmia electroanatomic mapping system (EAM).
Main Methods:
- Tested multiple connection configurations between DTA and Rhythmia EAM.
- Utilized Ensite EPT GenConnect box (GCB) and Rhythmia Maestro GCB (native setup).
- Evaluated alternative Medtronic GCB-E configurations.
Main Results:
- Real-time tracking of DTA on Rhythmia EAM was achieved in all tested setups.
- Configuration 3.1 showed an initial shift after the first radiofrequency (RF) application.
- Configuration 3.2 demonstrated no significant shift, while configuration 3.3 showed a massive positional shift.
Conclusions:
- A universal and reproducible compatibility testing method was established.
- DTA demonstrated satisfactory compatibility with Rhythmia EAM under a specific configuration.
- The findings highlight the importance of specific setup configurations for device compatibility.