Using Auditory Characteristics to Select Hearing Aid Compression Speeds for Presbycusic Patients
Yi Zhang1, Jing Chen2, Yanmei Zhang3
1Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Frontiers in Aging Neuroscience
|July 18, 2022
Summary
Selecting optimal hearing aid compression speeds for presbycusic patients can be guided by auditory characteristics. Action potentials (AP) and unaided modulation thresholds, along with speech recognition, help identify the best fast-acting compression (FAC) or slow-acting compression (SAC) settings.
Area of Science:
- Audiology
- Hearing Science
- Biomedical Engineering
Background:
- Presbycusis, or age-related hearing loss, affects temporal processing and speech-in-noise perception.
- Hearing aid compression speeds, fast-acting compression (FAC) and slow-acting compression (SAC), influence auditory signal processing.
- Optimal selection of compression speeds is crucial for improving hearing aid efficacy in presbycusic individuals.
Purpose of the Study:
- To determine the optimal hearing aid compression speeds (fast-acting and slow-acting) for presbycusic patients.
- To evaluate the impact of compression speeds on temporal modulation perception and speech-in-noise performance.
- To identify auditory characteristics that predict the best compression speed for individual patients.
Main Methods:
- 24 patients with moderate sensorineural hearing loss participated.
- Electrocochleogram (ECochG) including action potentials (AP) and summating potentials (SP) were recorded.
- Temporal modulation thresholds and speech recognition in noise were measured under unaided, FAC, and SAC conditions.
Main Results:
- Slow-acting compression (SAC) better preserved modulation thresholds compared to fast-acting compression (FAC).
- Speech recognition thresholds (SRT75%) did not significantly differ between FAC and SAC conditions (p=0.12).
- A decision tree model using AP, unaided modulation thresholds, and unaided SRT75% predicted optimal compression speed with 90% accuracy.
Conclusions:
- Both FAC and SAC improved speech recognition in noise for presbycusic patients.
- SAC may be superior in preserving modulation thresholds compared to FAC.
- AP, unaided modulation thresholds, and unaided SRT75% can guide personalized hearing aid compression speed selection.
More Related Videos
Related Concept Videos
Pulse amplitude and quality
2.1K
Pulse amplitude is a crucial indicator of cardiac health because it provides valuable insights into the strength of left ventricular contractions and the overall uniformity of blood circulation within the vasculature. The strength of the pulse is directly related to the force with which the heart contracts and the volume of blood being pumped.
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
2.1K
Perception of Sound Waves
4.6K
The human ear is not equally sensitive to all frequencies in the audible range. It may perceive sound waves with the same pressure but different frequencies as having different loudness. Moreover, the perception of sound waves depends on the health of an individual's ears, which decays with age. The health of one's ears may also be affected by regular exposure to loud noises.
The pitch of a sound depends on the frequency and the pressure amplitude of the source. Two sounds of the same...
The pitch of a sound depends on the frequency and the pressure amplitude of the source. Two sounds of the same...
4.6K
Hearing
53.0K
When we hear a sound, our nervous system is detecting sound waves—pressure waves of mechanical energy traveling through a medium. The frequency of the wave is perceived as pitch, while the amplitude is perceived as loudness.
53.0K