Wireless Communication Between Paired Leadless Pacemakers for Dual-Chamber Synchrony
Daniel J Cantillon1, Alok Gambhir2, Rajesh Banker3
1Cleveland Clinic, Heart and Vascular Institute, OH (D.J.C.).
Background:
Leadless pacemakers (LPs) can mitigate conventional pacemaker complications related to the transvenous leads and subcutaneous pocket surrounding the pulse generator. Although single-chamber leadless pacing has been established, multichamber pacing requires wireless bidirectional communication across multiple LPs to maintain synchrony. This preclinical study demonstrates the chronic performance of implant-to-implant (i2i) communication that achieves synchronous, dual-chamber pacing with 2 LPs.
Methods:
The i2i communication modality employs subthreshold electrical signals conducted between implanted LPs through the blood and myocardial tissue on a beat-by-beat basis. Right atrial and right ventricular LPs were implanted in 9 ovine subjects. The i2i transmission performance was evaluated 13 weeks after implant.
Results:
Right atrial and right ventricular LPs were implanted successfully and without complication in 9 ovine subjects. A total of 8715±457 right atrial-to-right ventricular and right ventricular-to-right atrial transmissions were sent per hour, with a success rate of 99.2±0.9%. Of periods with i2i communication failure when DDD pacing was not possible, 97.3±1.8% were resolved within 6 s.
Conclusions:
For the first time, synchronized, dual-chamber pacing has been demonstrated in a chronic preclinical feasibility study by 2 leadless pacemakers using beat-to-beat, wireless communication, achieving a success rate of 99.2%.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Phase-lead and Phase-lag Controllers
Mechanism of Cardiac Arrhythmias
Antiarrhythmic Drugs: Class I Agents as Sodium Channel Blockers
Class 1A Antiarrhythmic Drugs: These drugs work by moderately blocking sodium channels,...
Electrophysiology of Normal Cardiac Rhythm