Emmanuel J Favaloro1,2,3, Soma Mohammed1, Ronny Vong1
1Haematology, Institute of Clinical Pathology and Medical Research (ICPMR), NSW Health Pathology, Westmead Hospital.
Summary
This study harmonized coagulation factor testing procedures and normal reference ranges (NRRs) across a large network using ACL TOP 50 instruments. Results show comparable performance and acceptable factor sensitivity, enabling consistent laboratory diagnostics.
Area of Science:
- Clinical Chemistry
- Haemostasis Diagnostics
- Laboratory Medicine
Background:
- Coagulation factor testing is crucial for diagnosing bleeding disorders and investigating prolonged clotting times.
- Standardization of testing procedures and normal reference ranges (NRRs) is essential for accurate haemostasis laboratory results.
- Harmonization is needed across laboratory networks to ensure consistent patient care.
Purpose of the Study:
- To harmonize procedures and establish normal reference ranges (NRRs) for coagulation factor testing.
- To evaluate the performance of ACL TOP 50 family instruments and HemosIL reagents in a large laboratory network.
- To assess the comparability and factor sensitivity of new instrumentation against established methods.
Main Methods:
- Comparative evaluation of ACL TOP 550/750 instruments with HemosIL reagents against existing Stago and Siemens systems.
- Assessment of coagulation factor sensitivity in prothrombin time (PT) and activated partial thromboplastin time (APTT) assays.
- Establishment of coagulation factor NRRs using normal plasma samples.
Main Results:
- HemosIL factor assays demonstrated good comparability with reference methods (correlation coefficient R > 0.9).
- Coagulation factor sensitivity for PT and APTT assays was found to be acceptable, around 30 U/dl.
- Harmonized NRRs were successfully established across the laboratory network.
Conclusions:
- The ACL TOP 50 Family instruments and Werfen reagents provide acceptable performance for coagulation factor testing.
- The evaluation enabled the harmonization of coagulation factor testing procedures and NRRs within a large laboratory network.
- Standardized testing improves diagnostic accuracy and consistency for bleeding disorder assessment.