Catheters and dose optimization using a modified CVT algorithm and multi-criteria optimization in prostate HDR
Philippe Y Chatigny1,2, Cédric Bélanger1,2, Éric Poulin2
1Département de physique, de génie physique et d'optique, et Centre de recherche sur le cancer, Université Laval, Québec, Québec, Canada.
This study developed a new algorithm for optimizing catheter placement and dose in high-dose rate brachytherapy, potentially reducing treatment toxicity and the number of catheters needed for effective prostate cancer treatment.
Area of Science:
- Medical Physics
- Radiation Oncology
- Computational Biology
Background:
- Current high-dose rate (HDR) brachytherapy planning relies on planner experience for catheter positioning.
- Optimizing catheter number and location is crucial for dose coverage and reducing treatment toxicity.
- Patient-specific algorithms for catheter and dose optimization are needed.
Purpose of the Study:
- To develop and evaluate patient-specific algorithms for optimizing catheter placement and dwell times in HDR brachytherapy.
- To investigate the potential for reducing the number of catheters required while maintaining dose coverage and plan robustness.
- To assess the impact of catheter placement deviations on treatment plan quality.
Main Methods:
- A modified Centroidal Voronoi tessellation (CVT) algorithm was developed to optimize catheter positions.
- A graphics processing unit (GPU)-based multi-criteria optimization algorithm (gMCO) was used to optimize dwell times and positions.
- The algorithms were evaluated on 108 HDR brachytherapy cases for prostate cancer using real-time ultrasound, assessing plan robustness against insertion and reconstruction deviations.
Main Results:
- The combined CVT and gMCO approach increased the acceptance rate for Radiation Therapy Oncology Group (RTOG) criteria by 37% compared to standard methods.
- All patients met RTOG criteria with 11 catheters using CVT+gMCO with a 5 mm template, and 10 catheters with freehand placement.
- Plan acceptance remained high (85%) even with 3 mm deviations when using 11 catheters, with minimal decrease for 13+ catheters.
Conclusions:
- The developed CVT and gMCO algorithms enable optimization of catheter placement and dwell times for HDR brachytherapy.
- This approach can reduce the number of catheters needed for effective treatment while improving plan quality and robustness.
- Increased catheter numbers enhance plan robustness against insertion and reconstruction deviations.