Related Experiment Video
Updated: Sep 3, 2025

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Iatrogenic cardiac perforation due to pacemaker and defibrillator leads: a contemporary multicentre experience
Peter H Waddingham1,2, James Elliott1, Alexander Bates3
1Barts Heart Centre, St Bartholomew's Hospital, Barts Health NHS Trust, West Smithfield, London, EC1A 7BE, UK.
Aims:
To determine the incidence, clinical features, management, and outcomes of pacemaker (PM) and implantable cardioverter-defibrillator (ICD) lead cardiac perforation. Cardiac perforations due to PM and ICD leads are rare but serious complications. Clinical features vary widely and may cause diagnostic delay. Management strategies are non-guideline based due to paucity of data.
Methods And Results:
A multicentre retrospective series including 3 UK cardiac tertiary centres from 2016 to 2020. Patient, device, and lead characteristics were obtained including 6-month outcomes. Seventy cases of perforation were identified from 10 631 procedures; perforation rate was 0.50% for local implants. Thirty-nine (56%) patients were female, mean ( ± standard deviation) age 74 ( ± 13.8) years. Left ventricular ejection fraction 51 ( ± 13.2) %. Median time to diagnosis was 9 (range: 0-989) days. Computed tomography (CT) diagnosed perforation with 97% sensitivity. Lead parameter abnormalities were present in 86% (whole cohort) and 98.6% for perforations diagnosed >24 h. Chest pain was the commonest symptom, present in 46%. The management strategy was percutaneous in 98.6% with complete procedural success in 98.6%. Pericardial effusion with tamponade was present in 17% and was associated with significantly increased mortality and major complications. Anticoagulation status was associated with tamponade by multivariate analysis (odds ratio 21.7, 95% confidence interval: 1.7-275.5, P = 0.018).
Conclusions:
Perforation was rare (0.50%) and managed successfully by a percutaneous strategy with good outcomes. Tamponade was associated with increased mortality and major complications. Anticoagulation status was an independent predictor of tamponade. Case complexity is highly variable and requires skilled operators with a multi-disciplinary approach to achieve good outcomes.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
Related Concept Videos
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus....
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations:
Pericarditis I: Introduction
Cardiac Catheterization II: Right Heart Catheterization
Pericarditis III: Medical Management
Aneurysm III: Interprofessional Care