Biometry and intraocular power calculation using a swept-source optical coherence tomography: A repeatability and
Hassan Hashemi1, Mohammad Miraftab1, Parsa Panahi2
1Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran.
Purpose:
To evaluate the repeatability of biometry and intraocular lens (IOL) power using Galilei G6 and to determine the agreement of its measurements with those of IOL Master 700 and IOL Master 500.
Methods:
Hundred mature cataract eyes were examined twice with Galilei G6 and the results were compared with those of other two devices. Axial length (AL), minimum (K1), maximum (K2), and mean keratometry, anterior chamber depth (ACD), white-to-white (WTW) diameter, lens thickness (LT), and the calculated IOL power were the studied parameters. The correlation coefficient, within-subject standard deviation (Sw), Bland-Altman method, and 95% limits of agreement (LoA) were used for statistical analysis.
Results:
The intraclass correlation coefficient (ICC) was above 0.9 for all indices, and the LoA ranged from a minimum of 0.08 mm for AL to a maximum of 0.50 D for K1. Sw also ranged between a minimum of 0.02 for AL, ACD, and WTW and a maximum of 0.13 for K1. In the Galilei G6-IOL Master 700 pair, the narrowest and widest LoA were calculated for AL (0.07 mm) and K2 (0.49 D), respectively. In the Galilei G6-IOL Master 500 pair, the narrowest and widest widths of LoA were calculated for AL (0.17 mm) and K2 (0.92 D), respectively. In the first pair, the LoA of IOL power (0.57 D) were the best for Haigis formula and in the second pair, the best agreement (LoA: 0.35 D) was observed for Holladay-1.
Conclusion:
Galilei G6 provided repeatable biometric measurements. The agreement between biometry and IOL power calculation was better in the Galilei G6-IOL Master 700 pair compared to the Galilei G6-IOL Master 500.