Objective:
A prospective preoperative evaluation of 7 T ultra-high-field magnetic resonance imaging (MRI) in patients with suspected pituitary microadenomas for both adenoma detection and intrasellar localization compared with 3 T MRI was carried out.
Materials And Methods:
Patients underwent prospective preoperative standardized 3 and 7 T MRI. A distinct qualitative (lesion detection, intrasellar lesion location) and quantitative (lesion diameters, T1/T2 signal intensity ratio of the lesion to normal pituitary gland tissue) analysis was performed, along with an evaluation of image quality (IQ) regarding overall IQ, anatomical parameters, and artifacts; the findings of the qualitative analysis were compared with intraoperative findings and endocrinological outcomes.
Results:
Sixteen patients (mean age, 43 ± 16 years; 13 women) with pituitary microadenomas were included. Using 7 T MRI allowed the detection of 15 microadenomas-3 more than 3 T MRI. In addition, 7 T MRI allowed more precise lesion localization with 93.75% (15/16) agreement with intraoperative findings, compared with 75% (12/16) agreement using 3 T MRI. Lesion diameters showed no significant difference between 3 and 7 T MRI. T1 and T2 signal intensity ratio between microadenomas and normal pituitary gland tissue were higher in 7 T MRI than in 3 T MRI. The overall IQ and the IQ of each anatomical parameter of 7 T MRI were rated higher than those of 3 T MRI. No significant differences in susceptibility or head motion artifacts were observed between 3 and 7 T MRI; however, 7 T MRI was more susceptible to pulsation artifacts.
Conclusion:
Ultra-high-field MRI surpasses 3 T MRI in pituitary microadenoma detection and enables more precise delineation with higher correlation with intraoperative findings. Thus, 7 T sellar imaging is a promising option-especially in previously magnetic resonance-negative patients with endocrinologically confirmed hormone oversecretion-and helps reduce the need for invasive diagnostics.