Related Experiment Video
Updated: Sep 2, 2025

Hyperpolarized 13C Metabolic Magnetic Resonance Spectroscopy and Imaging
Published on: December 30, 2016
[Simulation Study on the Influence of Sampling Delay on the Accuracy of Energy Metabolism Measurement]
Anqi Zhang1,2, Yanyan Chen1, Yuan Wang1
1Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei, 230031.
Abstract:
Indirect energy metabolism measurement is the gold standard for providing nutritional support for critical illness. The accuracy of the measurement data directly affects the outcome of the disease. In order to study the influence of sampling delay on the accuracy of energy metabolism measurement under mechanical ventilation, the Matlab/Simulink platform and respiratory electrical model were used for simulation and quantitative analysis. The results show that the error of indirect energy metabolism measurement increases with the increase of sampling delay, the error of sampling delay in mechanical ventilation mode is larger than that of spontaneous breathing, and the error of sampling delay in PCV mode of mechanical ventilation is larger than that in VCV mode. Therefore, there should be different sampling delay compensation strategies under severe mechanical ventilation and its different control modes.