High-impedance alerts with pulse generator-Lead mismatch
Simon Christie1, Nada El Tobgy1, Colette M Seifer1
1Max Rady College of Medicine, Faculty of Health Sciences University of Manitoba Winnipeg Canada.
Background:
Cardiac Implantable Electronic Devices (CIED) include pulse generators and leads. In some implanting centers, it is a common practice to combine devices with leads from different companies. Case series have reported episodic high-impedance changes in Boston Scientific CIEDs with competitor leads. We investigated the incidence of high-impedance abnormalities in matched versus mismatched Boston Scientific Accolade pacemakers.
Methods:
A retrospective chart review identified all consecutive Boston Scientific Accolade pacemakers implanted between January 2017 and June 2019 at a Canadian tertiary care center. The primary outcome was the occurrence of transient, high-impedance changes which resulted in a switch to unipolar pacing/sensing in the absence of any other identifiable lead issue. Fisher exact tests (two-tailed, α = .05) were used to compare the incidence of outcomes in matched versus mismatched systems.
Results:
514 Boston Scientific Accolade pacemakers were associated with 882 individual leads. The primary outcome occurred with 21 leads (20 Medtronic and 1 Abbott), associated with occasional pacing inhibition, presyncope, and/or early surgical revision. Mismatched lead-device pairs were significantly associated with CIED malfunction compared to matched lead-device pairs (3.3% vs. 0%, p = .0019). The median time from implant to unipolar safety switch was 12.4 months. The median follow-up time was 21.6 months.
Conclusion:
Use of mismatched leads with a Boston Scientific Accolade device was associated with an increased risk of undesirable changes in sensing polarity with occasional inappropriate pacing inhibition. Awareness of this interaction can allow for the institution of appropriate programming remedies and may increase scrutiny of the use of mismatched CIED systems.
More Related Videos
07:17Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Impedance Combination
Line Protection with Impedance Relays
Under normal conditions, low load currents keep the measured...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...