Occurrence, diagnosis and treatment of thromboses of heart valve prostheses

Cor Et Vasa
|January 1, 1987
PubMed

Insights

Björk-Shiley mechanical heart valve thrombosis occurred at a rate of 0.36% per patient year, necessitating reoperation. Successful reoperations prevented recurrence, highlighting the importance of anticoagulation management.

Area of Science:

  • Cardiovascular Surgery
  • Biomaterials Science
  • Medical Device Engineering

Background:

  • Björk-Shiley mechanical heart valves were widely implanted between 1978 and 1985.
  • Thrombosis is a known complication associated with mechanical heart valves.
  • Different designs of Björk-Shiley valves exist, including flat disc and convexo-concave types.

Purpose of the Study:

  • To report the incidence of thrombosis in Björk-Shiley heart valve prostheses.
  • To evaluate the outcomes of reoperation for valve thrombosis.
  • To compare thrombosis rates between different Björk-Shiley valve designs.

Main Methods:

  • Review of 667 Björk-Shiley heart valve implantations from 1978-1985.
  • Analysis of 6 reoperations performed due to valve thrombosis.
  • Clinical and radiological assessment for diagnosis of thrombosis.

Main Results:

  • Six reoperations for thrombosis were performed, with a thrombosis rate of 0.36% per patient year.
  • All reoperations were successful, with no recurrence of thrombosis in the reoperated patients.
  • Thrombosis occurred exclusively in the flat disc Björk-Shiley valve type; no cases were observed in the convexo-concave type used since 1981.

Conclusions:

  • Strict anticoagulation management is crucial for preventing Björk-Shiley valve thrombosis.
  • Urgent reoperation is the preferred treatment for diagnosed disc valve thrombosis over thrombolysis.
  • The convexo-concave Björk-Shiley valve design may have a lower risk of thrombosis compared to the flat disc type.