Optimizing atrial sensing parameters in leadless pacemakers: Atrioventricular synchrony achievement in the real world
Sem Briongos-Figuero1, Álvaro Estévez-Paniagua1, Ana Sánchez Hernández1
1Cardiology Department, Infanta Leonor Hospital, Madrid, Spain; Complutense University, Madrid, Spain.
Background:
Performance of the leadless pacemaker capable of atrioventricular (AV) synchronous pacing in de novo patients warrants further investigation.
Objective:
The aims of this study were to assess what programming changes are needed to achieve proper atrial tracking and to study the percentage of AV synchrony (AVS) the device can provide under real-world conditions.
Methods:
Consecutive patients undergoing Micra AV implantation between June 2020 and November 2021 were studied. Reprogramming of atrial sensing parameters during follow-up was performed by following device counters. AVS was studied with an ambulatory 24-hour Holter monitor and automatically analyzed by an electrocardiogram delineation system. The primary end point was AVS ≥85% of total cardiac cycles during 24-hour Holter electrocardiogram monitoring.
Results:
Thirty-one patients who remained in VDD mode were studied, and all of them required manual reprogramming. The automatic A3 window end was deactivated, and a fixed and short value was set in all patients throughout follow-up. AVS significantly increased from 68.7% ± 14.7% at 24-hour follow-up to 83.9% ± 7.4% at 1-month visit (P = .001). At 1-month visit, shorter A3 window end time (P = .019), higher A4 threshold (P = .011), and deactivation of the automatic A3 window (P = .054) were independently related to higher AVS. A total of 2,291,953 Holter-recorded cardiac cycles were analyzed. Median AVS during 24-hour daily activities was 87.6% (interquartile range 84.5%-90.6%). Twenty of 26 patients (79.6%) reached AVS ≥85% of cardiac cycles.
Conclusion:
High rates of AVS can be achieved in real-world patients undergoing leadless pacing. Manual reprogramming of the atrial sensing parameters is essential to optimize mechanically sensed atrial tracking.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrophysiology of Normal Cardiac Rhythm
Mechanism of Cardiac Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias