The Relationship Between Different Bench Test Methodologies and Accuracy of Insulin Infusion Pumps: A Systematic
Fernanda Silva Tenorio1, Luiz Eduardo Galvão Martins2, Glória Sthefany Pereira da Silva2
1Escola Paulista de Medicina, Universidade Federal de São Paulo, São Paulo, Brazil.
Abstract:
Diabetes technology has rapidly evolved, and insulin infusion pumps (IIPs) have gained worldwide acceptance in diabetes care. The safety of medical equipment is highly discussed, imposing complex challenges in its use. The accuracy of IIPs can be determined through laboratory tests, generally following the IEC 60601-2-24 protocol. Studies have evaluated the accuracy and precision of IIPs, and there are discrepant results. So, we conducted a Systematic Literature Review to assess the methodologies used to evaluate the accuracy of IIPs, organizing the findings in a compiled perspective. The methodology was based on Kitchenham and Biolchini guidelines, and when possible it was carried out the Bayesian meta-analyses to compare the accuracy of IIPs. Most studies used the microgravimetric technique to evaluate the device accuracy, and some proposed adaptations for the standard protocol. The variation of results was recurrent, and the establishment of a protocol, especially to evaluate patch pumps, is necessary. The present study gives enough data to understand the scenario of the IIPs evaluation, as well as the different protocols that can be explored for its evaluation. This highlights the need for a reliable, practical, and low-cost methodology to assist the evaluation of IIPs.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Uncertainty in Measurement: Accuracy and Precision
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...