Possible limits of calibrating reading charts with the Landolt ring: a microscopic study
Wolfgang Radner1,2,3, Michael Radner4,5,6, Barbara Daxer4,5
1Karl Landsteiner University of Health Sciences, Dr. Karl-Dorrek-Straße 30, 3500, Krems, Austria. wolfgang.radner@chello.at.
Purpose:
To evaluate microscopically whether the print quality and accuracy of sizing of Landolt ring near vision charts are adequate for the calibration of reading charts.
Methods:
Near vision charts with Landolt rings from Oculus GmbH (C-Test; Wetzlar, Germany), Precision Vision (Woodstock, IL) and the RADNER Charts were examined, as well as custom-made Landolt rings optimized for print quality. Microscopic investigations and measurements were performed by using a Huvitz HSZ 600 stereomicroscope (Nikon NIS Elements software) to evaluate the height of the Landolt rings, the thickness of the lines, and the width of the openings. The deviations from the mathematically correct values, which were calculated as given in the EN/ISO 8596 and by the International Council of Ophthalmology (ICO), were analyzed (calculated for a test distance of 40 cm).
Results:
All the near vision charts showed notable deficiencies in print quality and aberrations from the nominal values in the height, thickness of the lines, and width of the openings. The openings were too narrow, whereas the height and thickness of the lines were larger than the nominal values. Even the openings of Landolt rings optimized for print quality were not always within an acceptable 5% tolerance and need further improvement.
Conclusion:
This study reports inaccuracies in the heights, thicknesses of the lines, and widths of the openings of Landolt rings in all the near vision charts investigated. The extent of these inaccuracies excludes such near vision charts as reference tests for the calibration of reading charts. The x-height in relation to the visual angle still seems to be the most reliable method for standardizing the print sizes for reading charts.
Related Concept Videos
Interpreting R Charts
An R chart plots the range of subsets of measurements collected from a process. Each point on the chart represents the range—defined as the difference between the maximum and minimum...
Uncertainty in Measurement: Reading Instruments
Calculations of Electric Potential I
The ring is divided into infinitesimal small arcs such that point M is equidistant from all the arcs. Here, the cylindrical coordinate system is used to calculate the electric potential at point M. A general element of the arc between angles θ and θ + dθ is of the...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
The R Chart
R charts are pivotal for pinpointing shifts in process variability. Stability is indicated when all data points remain within the defined upper and lower...