A High-Throughput Image-Guided Stereotactic Neuronavigation and Focused Ultrasound System for Blood-Brain Barrier Opening in Rodents
Published on: July 16, 2020
Phantom and Animal Study of a Robot-Assisted, CT-Guided Targeting System using Image-Only Navigation for Stereotactic
Khi Yung Fong1, Alexander Sheng Ming Tan2, Muhammad Syadad Bin Sulaiman3
1Yong Loo Lin School of Medicine, National University of Singapore, Singapore.
Purpose:
To evaluate the feasibility and accuracy of a robotic system to integrate and map computed tomography (CT) and robotic coordinates, followed by automatic trajectory execution by a robotic arm. The system was hypothesized to achieve a targeting error of <5 mm without significant influence from variations in angulation or depth.
Materials And Methods:
An experimental study was conducted using a robotic system (Automated Needle Targeting device for CT [ANT-C]) for needle insertions into a phantom model on both moving patient table and moving gantry CT scanners. Eight spherical markers were registered as targets for 90 insertions at different trajectories. After a single ANT-C registration, the closed-loop software targeted multiple markers via the insertion of robotically aligned 18-gauge needles. Accuracy (distance from the needle tip to the target) was assessed by postinsertion CT scans. Similar procedures were repeated to guide 10 needle insertions into a porcine lung. A regression analysis was performed to test the effect of needle angulation and insertion depth on the accuracy of insertion.
Results:
In the phantom model, all needle insertions (median trajectory depth, 64.8 mm; range, 46.1-153 mm) were successfully performed in single attempts. The overall accuracy was 1.36 mm ± 0.53, which did not differ between the 2 types of CT scanners (1.39 mm ± 0.54 [moving patient table CT] vs 1.33 mm ± 0.52 [moving gantry CT]; P = .54) and was not significantly affected by the needle angulation and insertion depth. The accuracy for the porcine model was 9.09 mm ± 4.21.
Conclusions:
Robot-assisted needle insertion using the ANT-C robotic device was feasible and accurate for targeting multiple markers in a phantom model.