Determination of polycyclic aromatic hydrocarbons in surface waters by high performance liquid chromatography
Emilia Frickel Critto1, Giselle Medina1, Mario Reta1
1Facultad de Ciencias Exactas, UNLP, CONICET, Laboratorio de Investigación y Desarrollo de Métodos Analíticos (LIDMA) y División Química Analítica, Calle 47 esq. 115, La Plata B1900AJL, Argentina.
Abstract:
A simple, sensitive and reproducible solid-phase extraction method using plastic cartridges containing a monolithic sorbent (m-SPE), coupled to reverse phase liquid chromatography analysis, aiming to determine fifteen polycyclic aromatic hydrocarbons in surface water samples, was developed. The sorbent was easily prepared through a thermal polymerization reaction by using a mixture of n-butyl methacrylate as non-polar monomer and ethylene glycol dimethacrylate as crosslinker contained in a typical Polypropylene syringe cartridge. The effect of different parameters (type of hydrophobic monomer, elution solvent, sample volume, sorbent amount and sorbent load capacity) on the extraction efficiency was optimized. The optimal conditions were achieved by using n-butyl methacrylate as monomer, tetrahydrofurane (THF) as solvent for sorbent cleaning, THF:acetone (1:1) as elution solvent, 25.00 mL of sample volume, 600 µL of the polymerization mixture and 60 µg L-1 as sample loading capacity. Finally, the sorbent charge capacity, the reusability of the cartridges and the extraction efficiency of the m-SPE monolith, as compared with a typical C8 cartridge, were evaluated. Under the optimized experimental conditions, enrichment factors were between 76 and 103, relative recovery factors from 78 to 103%, accuracy values in the range of 58 to 98%, and inter-batch reproducibility values from between 2 and 10%, were obtained. The limits of detection and quantification were obtained by two different procedures: the signal to noise (S/N) ratios (3 and 10, respectively) and the IUPAC convention. The lowest LOD and LOQ values, obtained with the S/N ratios, were between 0.02 and 1.00 µg L-1, respectively whereas with the IUPAC convention the values were between 0.07 and 5 µg L-1. Using this procedure, several PAHs could be detected in the surface water sample taken from a river stream located in La Plata city (Buenos Aires Province, Argentina).
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
10:34Characterization and Application of Passive Samplers for Monitoring of Pesticides in Water
Published on: August 3, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Types of Detectors